Dealer: This manual MUST be given to the end user.User: BEFORE using this product, read this manual and save for future reference.Operator’s Manual Fo
SECTION 1—GENERAL GUIDELINESPerfecto2™Series 10 Part No 1143482 WARNINGUse of some administration accessories or certain humidifiers not specified fo
SECTION 2—FEATURESPart No 1143482 11 Perfecto2™SeriesSECTION 2—FEATURESOxygen OutletFlowmeterOxygen Purity Indicator Lights / Fault and Power Indicato
SECTION 3—HANDLINGPerfecto2™Series 12 Part No 1143482SECTION 3—HANDLINGThe concentrator should ALWAYS be kept in the upright position to prevent cabin
SECTION 4—TYPICAL PRODUCT PARAMETERSPart No 1143482 13 Perfecto2™SeriesSECTION 4—TYPICAL PRODUCT PARAMETERSElectrical Requirements: 120 VAC ± 10% (132
SECTION 4—TYPICAL PRODUCT PARAMETERSPerfecto2™Series 14 Part No 1143482*NOTE: Stated concentration levels achieved after initial warm-up period (appro
SECTION 4—TYPICAL PRODUCT PARAMETERSPart No 1143482 15 Perfecto2™SeriesIRC5PO2 INDICATORSIRC5P INDICATORSLABEL SYMBOLO2 PURITY INDICATOR LIGHTS (LED)S
SECTION 5—OPERATING INSTRUCTIONSPerfecto2™Series 16 Part No 1143482SECTION 5—OPERATING INSTRUCTIONSIntroductionYour oxygen concentrator is intended fo
SECTION 5—OPERATING INSTRUCTIONSPart No 1143482 17 Perfecto2™SeriesRecommended Guidelines for Optimum PerformanceNOTE: Ensure that your concentrator i
SECTION 5—OPERATING INSTRUCTIONSPerfecto2™Series 18 Part No 1143482Connect Humidifier (If So Prescribed) WARNINGFor optimum performance, Invacare rec
SECTION 5—OPERATING INSTRUCTIONSPart No 1143482 19 Perfecto2™SeriesFIGURE 5.2 Humidifer Bottle Adapter5. Attach the humidifier bottle adapter to th
WARNINGPerfecto2™Series 2 Part No 1143482 WARNINGDO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE
SECTION 5—OPERATING INSTRUCTIONSPerfecto2™Series 20 Part No 1143482FIGURE 5.3 Humidifier Compartment
SECTION 5—OPERATING INSTRUCTIONSPart No 1143482 21 Perfecto2™SeriesPower SwitchNOTE: For this procedure, refer to FIGURE 5.4.1. Press power switch to
SECTION 5—OPERATING INSTRUCTIONSPerfecto2™Series 22 Part No 11434822. If the flowrate on the flowmeter ever falls below 0.5 L/min for Perfecto2 5 for
SECTION 5—OPERATING INSTRUCTIONSPart No 1143482 23 Perfecto2™SeriesInitial Startup of the ConcentratorNOTE: Concentrator may be used during the initia
SECTION 5—OPERATING INSTRUCTIONSPerfecto2™Series 24 Part No 1143482Indicators (If Your Unit Does Not Feature the O2 Sensor) NOTE: For this procedure,
SECTION 6—MAINTENANCEPart No 1143482 25 Perfecto2™SeriesSECTION 6—MAINTENANCENOTE: The Invacare concentrators are specifically designed to minimize ro
SECTION 6—MAINTENANCEPerfecto2™Series 26 Part No 1143482FIGURE 6.1 Cleaning the Cabinet FilterCabinet WARNINGUNPLUG the concentrator when cleaning.
SECTION 6—MAINTENANCEPart No 1143482 27 Perfecto2™SeriesPREVENTIVE MAINTENANCE RECORD Model No. IRC _______________________________Serial No._________
SECTION 7—TROUBLESHOOTING GUIDEPerfecto2™Series 28 Part No 1143482SECTION 7—TROUBLESHOOTING GUIDESYMPTOM: PROBABLE CAUSE: SOLUTION:Alarm: Main Power L
SECTION 7—TROUBLESHOOTING GUIDEPart No 1143482 29 Perfecto2™SeriesYELLOW or RED light Illuminated.Alarm:ContinuousOn RED light only.* Only applies to
TABLE OF CONTENTSPart No 1143482 3 Perfecto2™SeriesTable of ContentsACCESSORIES ... 2SPECIAL NOTES ...
SECTION 7—TROUBLESHOOTING GUIDEPerfecto2™Series 30 Part No 1143482Alarm:RapidBeep...Beep...Beep...BeepLOW FLOW ALARM1a. Kinked or blocked tubing, can
LIMITED WARRANTYPart No 1143482 31 Perfecto2™SeriesLIMITED WARRANTYNOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR S
Table of ContentsTABLE OF CONTENTSPerfecto2™Series 4 Part No 1143482Indicators (If Your Unit Does Not Feature the O2 Sensor) ...24SECTION 6—MAINTENANC
SPECIAL NOTESPart No 1143482 5 Perfecto2™SeriesSPECIAL NOTESSignal words are used in this manual and apply to hazards or unsafe practices which could
SECTION 1—GENERAL GUIDELINESPerfecto2™Series 6 Part No 1143482SECTION 1—GENERAL GUIDELINESIn order to ensure the safe installation, assembly and opera
SECTION 1—GENERAL GUIDELINESPart No 1143482 7 Perfecto2™Series DANGERThe use of oxygen therapy requires that special care be taken to reduce the risk
SECTION 1—GENERAL GUIDELINESPerfecto2™Series 8 Part No 1143482 WARNINGNEVER block the air openings of the product or place it on a soft surface, such
SECTION 1—GENERAL GUIDELINESPart No 1143482 9 Perfecto2™Series WARNINGTo Reduce The Risk Of Burns, Electrocution, Fire Or Injury To Persons.Avoid usi